I would like to use GuruxDirector to read data from STZ351 from star instrument via optical probe. do you have any default setting for this meter.
Thanks
Try to select Gurux as a manufacturer. Then set serial port settings. You can try o use IEC first, but if meter don't reply, change start protocol to DLMS and try with 9600 8N1.
Been a while. I finally manage to connect it with DLMS director by tying all the combination of setting. The setting is :
DLMS protocol
19200:8None1
Authentication : Low
Password : 12345678
Server address : 1 (Logical server 0, Physical server 1)
Client address : 1
Thank you wey much for this info. Shenzhen Star Instrument manufacturer settings are added, so it's easier to read in the future. Finding correct settings is really frustrating and it might take a very long time.